Model Number GIF-HQ190
Medical Device Problem Codes No Display/Image (1183); Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection, and the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause was traced to a component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
 
Event or Problem Description
The customer returned the gastrointestinal videoscope to the service center for a separate issue.During the initial evaluation of the returned device, a scope communication error e216 and no image were observed.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Common Device Name
Gastroscope And Accessories, Flexible/Rigid
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50923724
Report Number9610595-2026-79866
Device Sequence Number5753379
Product Code FDS
UDI-Device Identifier04953170305276
UDI-Public04953170305276
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K131780
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-HQ190
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Supplement Date Received by Manufacturer09/30/2026
Initial Report FDA Received Date09/28/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-