During a procedure, a cancellation was noted when the ensite x field frame was not being recognized by the system after connecting it.Disconnecting and reconnecting the field frame as well as restarting the ensite x amplifier resolved the issue.It was also noted that after connecting the catheters to the amplifier, there was an "amplifier error, cannot communicate with amplifier" message and the catheters were repeated shown connected and then disconnected.Trying the catheters on different ports and restarting the ensite x system di not resolve the issue and the catheters would only show up for a few seconds on the ensite.Mapping was not possible and when ending the study, an "unexpected error" popup was shown.Resuming the procedure after rebooting the ensite x dws was not possible, and the procedure could not be finished as planned.Further checking found that the field frame connections against the connector looked charred.There were no adverse patient consequences.
|