• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Beta Bionics, Inc. iLet Bionic Pancreas; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Beta Bionics, Inc. iLet Bionic Pancreas; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Display Difficult to Read (1181)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
iLet Bionic Pancreas
Common Device Name
ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP
Manufacturer (Section D)
Beta Bionics, Inc.
11 hughes
irvine CA 92618
Manufacturer (Section G)
Beta Bionics, Inc.
11 hughes
irvine CA 92618
Manufacturer Contact
jared fukushima
11 hughes
irvine, CA 92618
+1(949)288-8350
MDR Report Key50926734
Report Number3019004087-2026-68026
Device Sequence Number18523162
Product Code QFG
UDI-Device Identifier00850050080206
UDI-Public00850050080206
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K231485
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age28 YR
Patient SexFemale
-
-