| Model Number |
MAJ-210 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported during a bronchoscopy, while applying anesthetic via a syringe, a fragment of the device was found to have detached.The fragment entered the syringe, so there was no drop inside the body.The procedure was completed after replacement with the similar product.There were no reports of health hazard due to this event and no delay in reporting.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for inspection.The root cause could not be identified.Based on the results of the investigation, the customer allegation was confirmed the biopsy valve broke and fell off.It is likely the biopsy valve broke due to stress caused by inserting or removing the syringe at an angle resulting in component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|