U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant SINGLE USE BIOPSY VALVE (Sterile)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant SINGLE USE BIOPSY VALVE (Sterile) Back to Search Results
Model Number MAJ-210
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported during a bronchoscopy, while applying anesthetic via a syringe, a fragment of the device was found to have detached.The fragment entered the syringe, so there was no drop inside the body.The procedure was completed after replacement with the similar product.There were no reports of health hazard due to this event and no delay in reporting.
 
Additional Manufacturer Narrative
The device was returned for inspection.The root cause could not be identified.Based on the results of the investigation, the customer allegation was confirmed the biopsy valve broke and fell off.It is likely the biopsy valve broke due to stress caused by inserting or removing the syringe at an angle resulting in component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE BIOPSY VALVE (Sterile)
Common Device Name
SINGLE USE BIOPSY VALVE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant
34-3 hirai, hinode-machi
nishitama-gun, tokyo 190-0 182
JA  190-0182
Manufacturer (Section G)
OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant
34-3 hirai, hinode-machi
nishitama-gun, tokyo 190-0 182
JA   190-0182
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50927050
Report Number3003637092-2026-00191
Device Sequence Number16133686
Product Code EOQ
UDI-Device Identifier14953170152433
UDI-Public14953170152433
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K061313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-210
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/09/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3003637092-12/29/2025-00
Patient Sequence Number1
Patient SexUnknown
-
-