| Model Number |
IO-501 |
| Medical Device Problem Codes |
Failure to Run on Battery (1466); No Flow (2991)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Low Oxygen Saturation (2477)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's unit stopped working following a routine battery change.As a result, the patient's oxygen levels started to drop and they experienced shortness of breath.They were able to make it back home and use their stationary concentrator without further intervention.A replacement unit was sent to the patient and it is working for them.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is estimated.The unit was returned for evaluation on 16-sep-2026.Unit came in for oxygen error.Unit has no 02 errors.All numbers are within specification and there is no popped-up error on the data log, however, dirty inlet filter due to dust and lower housing mount damage due to compressor vibration.
|
| |
|