• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Inogen Inc. Inogen Rove 6 Portable Oxygen Concentrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Inogen Inc. Inogen Rove 6 Portable Oxygen Concentrator Back to Search Results
Model Number IO-501
Medical Device Problem Codes Failure to Run on Battery (1466); No Flow (2991)
Health Effect - Clinical Codes Dyspnea (1816); Low Oxygen Saturation (2477)
Date of Event 09/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's unit stopped working following a routine battery change.As a result, the patient's oxygen levels started to drop and they experienced shortness of breath.They were able to make it back home and use their stationary concentrator without further intervention.A replacement unit was sent to the patient and it is working for them.
 
Additional Manufacturer Narrative
B3: date of event is estimated.The unit was returned for evaluation on 16-sep-2026.Unit came in for oxygen error.Unit has no 02 errors.All numbers are within specification and there is no popped-up error on the data log, however, dirty inlet filter due to dust and lower housing mount damage due to compressor vibration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen Rove 6 Portable Oxygen Concentrator
Common Device Name
Inogen Rove 6 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50927274
Report Number3004672275-2026-00316
Device Sequence Number16133691
Product Code CAW
UDI-Device Identifier00817131020513
UDI-Public(01)00817131020513
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K230052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-501
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/16/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-