Medtronic Puerto Rico Villalba SENSIGHT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
| Model Number |
B3301542M |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that impedances were out of range when tested in the operating room, with red contacts noted on contacts 2c and 3 after the lead was placed.The hcp and manufacturer representative attempted multiple troubleshooting steps, including removing and wiping down the lead, but the impedances did not clear.The lead was removed and replaced with a second lead, which showed good impedances.The issue was resolved following surgical intervention, and no further information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|