Catalog Number DYNJRA2433
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that needle (component # (b)(4)) is not sharp enough and is a good bit duller than the previous needle (component # (b)(4)).
 
Additional Manufacturer Narrative
The account reported finding dull needle; according to the facility, during a pain management injection, the needle was observed to be more dull than expected, the needle was discarded and replaced, and no re-stick was required.A photo was received for investigation, and the customer provided a photo showing the provider holding two needles up; the only component of the needles that are visible are the needle hubs.The packs production record and bill of materials were reviewed, and no discrepancies were noted that would lead to the reported issue.Based on the complaint details and investigation, the complaint is not confirmed; the investigation was limited without receipt of the physical sample, and sample required to determine cause.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
PAIN TRAY B THOMPSON
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sullivan bluhm
three lakes drive
northfield, IL 60093--275
MDR Report Key50927325
Report Number1417592-2026-52808
Device Sequence Number10735216
Product Code OFT
UDI-Device Identifier10198459036248
UDI-Public10198459036248
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJRA2433
Device Lot Number25JBS009
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1