MEDTRONIC PUERTO RICO OPERATIONS CO., HU CD HORIZON® Spinal System; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
| Model Number |
7078396 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from the healthcare professional via a manufacturer representative regarding a device used for l3-il rear fixation + l5/s posterior lumbar interbody fusion and hernia lcs was the preoperative diagnosis of this surgery and there was a less than 60 mins delay in the overall procedure time.It was reported that all rods were installed and the final was performed, but during the final of the iliac, there was a strange noise and a bur appeared, so the final (thread cutting) could not be finished.Even after replacing the set screw, the same issue occurred, so a new screw was swapped in and the final was performed without any issues.There were no patient symptoms reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter is unknown/ first name and last name of initial reporter is unknown.G2: the country of the event is japan.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|