| Model Number |
27003 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During resmed evaluation, an astral device displayed an error message (sf140) related to alarm priority.There was no harm or serious injury reported as a result of the incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The main circuit board was replaced to address the issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).During a routine check of complaints records it was detected that the event is reportable in the us.This report is being submitted to correct the error.
|
| |
|