| Model Number |
480522-02 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the vessel sealer curved instrument jaws would not open.The procedure was completed.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the vessel sealer curved (vsc) instrument to perform failure analysis (fa).Fa was able to confirm the reported complaint.The instrument was found to have the grip ring dislodged at the proximal end (in the housing).As a result, the instrument grips would not open when driven on the system or when the grip open lever was used.The grip ring set screw was found rattling inside of the housing.Additionally, the vsc instrument exhibited imprecise motion during grip open function.The grip tips moved within the joint limits and in the commanded direction; however, the grips would not open.The instrument was found to have the grip ring dislodged which prevented the grips from opening as expected.The probable root cause is attributed to the observed dislodged instrument grip ring not up against lever during the set grip ring position process step.
|
| |
|