Model Number 480522-02
Medical Device Problem Codes Break (1069); Failure to Seal (4070)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the sealing was not completed when using a vessel sealer curved instrument.No known impact or patient consequence was reported.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive the vessel sealer curved instrument to perform failure analysis (fa).Fa was not able to confirm nor reproduce the customer reported complaint.The instrument was tested on an in-house system, and it passed the self-tests.The instrument moved intuitively with full range of motion in all directions and the grip tips opened and closed properly. the vessel sealer curved instrument passed energy delivery testing in various grip orientations and also passed the electrical continuity test and the cut test.A review of the logs found that the instrument was used for a total of 440 seconds.During that time the instrument generated two activation errors.Additional evaluation found the vessel sealer curved instrument had the lower grip tip broken at the distal end.The broken piece measured approximately 0.87mm x 1.75mm and was not returned with the instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
da Vinci Energy
Common Device Name
Vessel Sealer Curved
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50927510
Report Number2955842-2026-40140
Device Sequence Number10759105
Product Code NAY
UDI-Device Identifier10886874123369
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K250674
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number480522-02
Device Catalogue Number480522
Device Lot NumberU10251011 0028
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/17/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-