Effective august 1, 2026, mdr-related activities and responsibilities were transferred to pentax medical corporation as part of a corporate restructuring.This report is being submitted by pentax medical corporation.In accordance with 21 cfr 803.3(m), because this corporate restructuring does not affect the manufacturing site, the fda registration number previously assigned for this site (#9610877) continues to be used.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the insertion flexible tube steel braid piercing.Based on the result, we concluded that it was caused due to the physical damage applied on the insertion flexible tube.In addition, our technician confirmed that the lcb distal cover glass cracked, the air nozzle clogged, the insertion flexible tube leak, and the suction channel kink; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0628 (ift)" and/or the risk analysis results, it was evaluated to submit mdr.
|