| Model Number |
TJF-Q190V |
| Medical Device Problem Codes |
Crack (1135); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the duodenovideoscope to the service center for a separate issue.During initial evaluation of the device, the objective lens and the light guide lens had chipped glue and chipped lenses.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the final investigation of the device the initial evaluation findings were confirmed.Based on the results of the investigation, a likely cause of malfunction is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|