Updated fields:b1,b4,b7,d8,g3,g6,g7,h2,h6(type of investigation, investigation findings,component codes, investigation conclusions),h11
corrected fields: d4(udi),d10,h4
additional contact info:(b)(6)
please revert the a1(patient id),d8 field to blank
product was not returned so an exact cause of the issue could not be identified.However, based on our investigation, difficult or unable to insert through sheath occur due to one or more of the following probable causes.The failure mode difficult or unable to insert iab through sheath occurs when the intra aortic balloon catheter cannot be smoothly advanced through the supplied introducer sheath.This may involve increased resistance, halted progression, or complete blockage.The issue can compromise appropriate iab placement, delay therapy, or require device replacement.Potential causes
insertion difficulty can result from abnormalities in either the sheath or catheter, including
sheath related causes
kinked or bent sheath
unfolded sheath
clotted sheath
incorrect sheath size
damaged sheath hub
catheter lumen related causes
broken inner lumen
inner lumen kink
catheter kink.Unfolded membrane.
|