• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. MEGA 8Fr. 50cc IAB; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. MEGA 8Fr. 50cc IAB; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0498-01
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields:b1,b4,b7,d8,g3,g6,g7,h2,h6(type of investigation, investigation findings,component codes, investigation conclusions),h11 corrected fields: d4(udi),d10,h4 additional contact info:(b)(6) please revert the a1(patient id),d8 field to blank product was not returned so an exact cause of the issue could not be identified.However, based on our investigation, difficult or unable to insert through sheath occur due to one or more of the following probable causes.The failure mode difficult or unable to insert iab through sheath occurs when the intra aortic balloon catheter cannot be smoothly advanced through the supplied introducer sheath.This may involve increased resistance, halted progression, or complete blockage.The issue can compromise appropriate iab placement, delay therapy, or require device replacement.Potential causes insertion difficulty can result from abnormalities in either the sheath or catheter, including sheath related causes kinked or bent sheath unfolded sheath clotted sheath incorrect sheath size damaged sheath hub catheter lumen related causes broken inner lumen inner lumen kink catheter kink.Unfolded membrane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 8Fr. 50cc IAB
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50927781
Report Number2248146-2026-0001576
Device Sequence Number21522438
Product Code DSP
UDI-Device Identifier10607567107301
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0498-01
Device Catalogue Number0684-00-0498-01
Device Lot Number3000469255
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IAB.
Patient Age60 YR
Patient SexMale
Patient Weight143 LB
-
-