| Catalog Number |
CQF7594 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a 64-year-old male patient was prepared for a percutaneous transluminal angioplasty (pta) procedure using a conquest 40 balloon catheter via a right arm arteriovenous fistula.Prior to patient use, during device preparation, the conquest balloon catheter was advanced through the sheath; however, the balloon failed to inflate, and fluid leakage was observed from the hub due to a crack.It was further reported that a complete fracture of the balloon catheter was identified.The device was not used on the patient.The procedure was subsequently completed using another balloon catheter, resulting in a procedural delay.There was no reported patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported detachment could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D2b (dqy; lit), h.4.Device manufacture date: unknown.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|