The customer reported the sphygmomanometer will not inflate.Medline received the sample in like-new condition, performed a visual inspection, and tested the sample according to the instructions for use; the cuff was inflated five times up to 300 mmhg, and the gauge displayed measurements without any issues.The complaint was not confirmed, and no manufacturing defects were noticed that would prevent the sample from working.The sample operated as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|