| Model Number |
SDM |
| Medical Device Problem Code |
Unintended Electrical Shock (4018)
|
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An incident occurred with sentec sdm device (sn: (b)(6) in which one of the employees of the munich firebrigade (germany) received an electric shock.The device was mounted on a stretcher together with an incubator and other associated equipment.The monitor was to be detached from the stretcher.While doing so, the employee touched the rear of the device and received an electric shock.A minor burn mark was visible on her hand, and both the residual current device (rcd) of the stretcher and the rcd of the station tripped.The device was then disconnected from the mains and taken out of service.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently in transit back to the manufacturer for root cause analysis.
|
| |
|