Medline Industries, LP MEDLINE; GLOVE,EXAM,PF,NITRILE,SMARTGUARD,SMALL Back to Search Results
Catalog Number SG311
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "staff are saying they look the same, feel and fit different, and are tearing while applying.".
 
Additional Manufacturer Narrative
The reporter described issues with smart guard exam gloves ripping and stated that staff were saying the gloves look the same, feel and fit different, and are tearing while applying.Medline received samples and was able to observe the reported issue of tear via the supplied image; from the provided 2 samples for evaluation, both samples had acceptable elongation and did not tear prematurely while donning, and after comparing the feel to other gloves from the same manufacturer, there was no noticeable difference.The reported issue was confirmed via the supplied images, but the provided samples did not tear prematurely while donning.The cause code was updated to, supplier error asia (medline brand) machine/automation process error.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
GLOVE,EXAM,PF,NITRILE,SMARTGUARD,SMALL
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
herchelle teodoro
three lakes drive
northfield, IL 60093--275
MDR Report Key50927984
Report Number1417592-2026-52818
Device Sequence Number9734263
Product Code LZA
UDI-Device Identifier00884389172028
UDI-Public00884389172028
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K051378
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSG311
Device Lot NumberME401887194
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/20/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2024
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-