It was reported that during a training/demo of the da vinci system, the clinical sales representative (csr) observed the right master tool manipulator (mtm) was not working properly on the system.A work order was requested for demo/not for human use (nfhu) system due to the right mtm malfunction.Logs were not available at the time of the report.Following a restart of the system, the right mtm remained unstable.It was also noted that ergonomics needed to be checked, though no further information was provided on that matter.There was no report of patient involvement or reported injury.Intuitive surgical, inc.(isi) obtained the following additional information: on this console, none of the ergonomic adjustments were working.The armrest height, pedal depth, and viewer were completely non functional, and the message ¿ergonomics error¿ appeared on the display.The console was restarted several times, and the ergonomics only started working after approximately the fifth restart.However, this did not last long, and the right mtm stopped working during a simulation exercise.The console was restarted several times.In this condition, the console is not suitable for demonstration purposes.
|