| Model Number |
DDU-120E |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An international distributor reported that their customer's aed asi is flashing red, and the device is reporting a service code.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Although the customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting within the unit.Analysis of the aed's log files identified that once a new battery was installed and a manual self test run the aed functioned as designed and could stay in service.Analysis of the aed's log files did not identify a cause for the depleted battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.
|
| |
|