B. Braun Melsungen AG EASYPUMP; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 4540008-02
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A review of the batch manufacturing record (bmr) was performed for the reported lot number, and no abnormalities or non-conformances were noted during the in process or final product inspection.Final control flow rate report of the affected batch 26a09ge511 was reviewed.For final control flow rate report, the average flow rate deviation from the nominal flow rate was between 1.08% and 5.83%.As no returned sample was received, further investigation was not possible.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP
Common Device Name
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
B. Braun Melsungen AG
carl-braun strabe 1
melsungen, hessen 34212
GM  34212
Manufacturer (Section G)
B. Braun Melsungen AG
carl-braun strabe 1
melsungen, hessen 34212
GM   34212
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
+1(833)425-1464
MDR Report Key50928002
Report Number9610825-2026-00524
Device Sequence Number21522313
Product Code MEB
UDI-Device Identifier04046964448485
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K081905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4540008-02
Device Lot Number26A09GE511
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-