| Model Number |
380652-50 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted general inguinal hernia surgical procedure, a recoverable fault 32056 had been reported and error 32056 on universal surgical manipulator (usm) arm 4 was observed.Technical service engineer (tse) confirmed the error on the system logs.The customer performed an emergency power off (epo) of the patient side cart (psc); however, the issue persisted.The procedure was continuing with usm arm disabled with no report of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse replaced the universal surgical manipulator (usm).The system was tested and verified as ready for use.Isi did receive a da vinci product involved with this complaint to perform failure analysis.However, failure analysis is not complete.
|
| |
|