It was reported that during a da vinci-assisted surgical procedure, the customer encountered a repeating error 32098 on the universal surgical manipulator 3 (usm3).The customer was able to recover the error by adjusting the usm position but the error would return.The technical service engineer advised the customer to perform a hard power cycle when possible.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the customer did disabled the arm 3 and completed the case with the remaining three arms and a laparoscopic instrument in the arm 3 port.The issue cased a 30 minute surgical delay.
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the universal surgical manipulator (usm) 3.The system was verified and ready for use.Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.Blank mdr fields:
the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Field e4 is blank because it is unknown if the initial reporter submitted a report to the fda.Fields g5 and g7 are not applicable.Field h10 is blank as there are no related report numbers.
|