| Model Number |
380652-55 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted radical salvage prostatectomy surgical procedure, the site stated they were having recurring recoverable error 32056 on universal surgical manipulators (usm) 1.The nurse restarted multiple times and error returned.The nurse was able to usm 1.The technical support engineer (tse) reviewed logs and informed that usm 1 was faulty and it has to be replaced.Upon request, the nurse was not able to state if the procedure could take place with the 3 remaining arms.The surgeon took over the call and did not answer to the question.Tse called back reporter to inform about the next step and to collect procedure information.The site informed that they would proceed with 3 universal surgical manipulators (usm).The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the universal surgical manipulator (usm) 1.The system was verified and ready for use.Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.
|
| |
|