| Catalog Number |
733010E-B |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
-customer reported that during preparation of a 10x30mm straight graft, white particulate/flaking material was observed coming from the inside of the graft.The customer described the material as white flakes that were thicker than a powder.When the graft was manipulated or "wiggled," the material continued to flake off and was visible on the clinician's glove.Due to the amount of particulate observed, the first graft was passed off the field and is being returned for analysis.( reported under manufacturing report number: 9612515-2026-00062).-a second 10x30mm graft form the same lot was opened.The customer reported that the second graft demonstrated similar white flaking, although to a much lesser extent than the first graft this was implanted successfully with no issues.( this is being reported under manufacturing number 9612515-2026-00063).
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer narrative:
clinical code: e
4582 - no clinical signs, symptoms or conditions - device implanted , outcome of procedure was good.Impact code: f
2199- no health consequences or impact - this second device was noted to have white particulate/flaking material however to a lesser extent then the original ( reported under report number: 9612515-2026-00062) this graft was implanted successfully outcome of procedure was good.Device problem code: a
1261 - material fragmentation- white particulate/flaking material was observed coming from the graft.The customer described the material as white flakes that were thicker than a powder.Component code: g
4755 - part/component/sub-assembly term not applicable.Type of investigation: b
4109 - historical data analysis: - gelweave sales jan 22 - aug 26 vs gelweave gelatin detached / flaking events jan 22 - sep 26 gave an occurrence rate of (b)(4) no trend requiring action has been identified.4112 - analysis of information provided by user/third party: additional information has been requested to aid this investigation.3331 - analysis of production records: full batch review performed from base fabric to finished product for device id- 26370041 which confirmed the batch was manufactured to specification.4117 - device not accessible for testing - device remains implanted.Investigation findings: c
3233 - results pending completion of investigation.Investigation conclusion: d
11 - conclusion not yet available as investigation is ongoing.
|
| |
|
Search Alerts/Recalls
|