It was reported that during a da vinci-assisted surgical procedure, a remote site representative had contacted the intuitive surgical, inc.(isi) technical support engineer (tse) on behalf of the user site to report that monopolar energy could not be activated and that the cord indicator on the energy switch module had been amber while the other indicators had been blue.Prior to the call, users had verified that all cords had been connected correctly and had replaced the energy cord and the instrument, but the issue had persisted.The tse had asked the caller to verify that the instrument cord had been connected to the energy switch module rather than the generator, and the caller had confirmed this had already been verified based on a photo of the setup.The tse had then advised restarting the system, and the caller had planned to relay that instruction to the user site and call back; the call had ended before additional troubleshooting details had been obtained.The procedure outcome is unknown with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: the case was completed robotically but the monopolar energy wasn't work at all.The customer tried the cable changing, emergency power off (epo), everything but nothing was changed.The surgeon cut the tissues without using energy.There was a such a big delay of course maybe more than 60min.The total operation time was almost 5hours.At the end of the surgery, the monopolar energy suddenly worked.After the surgery their technical support came to fix the system but the energy was working correctly; so they could not do anything.
|