| Model Number |
381121-44 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted pulmonary lobectom surgical procedure, user informed the technical support engineer (tse) that erbe error c-34 occurred repeatedly.The user stated the erbe generator had already been restarted, but the issue persisted.The user continued with the procedure using the ft10 generator without further issues.No known impact or patient consequence was reported.Intuitive followed up but no additional information was available.No patient information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse replaced the integrated electrosurgical unit (iesu).The system was tested and verified as ready for use.Isi did not receive a da vinci product involved with this complaint to perform failure analysis.
|
| |
|