• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC daVinci Xi; PATIENT SIDE CART, 4-ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC daVinci Xi; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380652-56
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that outside of da vinci-assisted surgical procedure, site contacted intuitive technical support engineer (tse) to report crack in carriage of universal surgical manipulator (usm) 2.There was no report of patient involvement.
 
Additional Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the universal surgical manipulator (usm) 2.The system was verified and ready for use.Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
daVinci Xi
Common Device Name
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50928035
Report Number2955842-2026-40278
Device Sequence Number16133697
Product Code NAY
UDI-Device Identifier00886874110720
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number380652-56
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient SexUnknown
-
-