| Model Number |
CI622 |
| Medical Device Problem Code |
Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Per the clinic, the device was explanted on (b)(6) 2026 due to the poor sound quality.There are no plans to re-implant the patient with a new device as of the date of this report.Additional information has been requested but has not been made available as of the date of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 event site name: st vincent's private hosp northside
a getinge field service engineer (fse) was dispatched to evaluate the unit.The fse tested the compressor reading and front fan connection.The fse tested the device but after 45 minutes the hissing noise started again.The fse replaced the muffler as found faulty.Performed the full testing and calibrations as per specifications.The unit passed all safety and functional tests and was working within specifications.
|
| |
|