• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (Boxborough) O-Arm O2 Imaging System; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (Boxborough) O-Arm O2 Imaging System; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding an imaging system being used in an unknown procedure.It was reported that there was a navigation inaccuracy in connection with the imaging system.Patient presence was unknown.
 
Additional Manufacturer Narrative
H3) the manufacturer representative went to the site to test the imaging system.The site wanted a system checkout.The camera temporarily recognized the reference frame or navlock tracker orange.The navigation system had displayed the message localizer faulty.Furthermore, a navigation inaccuracy was reported in connection with imaging system.Function check carried out, camera, camera cable and connectors checked for contact problems.No error was found.Checked navigation accuracy in conjunction with the imaging system.The imaging system tracker calibration verified and passed on both sides.Navigation system in conjunction with the imaging system was ready for operation.Ndi event log checked.Warning message: "firmware incompatibility detected".After restart system info message: "firmware compatibility restored".Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi-500-01145, serial/lot #: 4.2.0, ubd: , udi#:.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-Arm O2 Imaging System
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough MA 01460
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50928083
Report Number3004785967-2026-00521
Device Sequence Number18825147
Product Code OWB
UDI-Device Identifier00643169618121
UDI-Public(01)00643169618121(11)161129
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11...."
Patient SexUnknown
-
-