H3) the manufacturer representative went to the site to test the imaging system.The site wanted a system checkout.The camera temporarily recognized the reference frame or navlock tracker orange.The navigation system had displayed the message localizer faulty.Furthermore, a navigation inaccuracy was reported in connection with imaging system.Function check carried out, camera, camera cable and connectors checked for contact problems.No error was found.Checked navigation accuracy in conjunction with the imaging system.The imaging system tracker calibration verified and passed on both sides.Navigation system in conjunction with the imaging system was ready for operation.Ndi event log checked.Warning message: "firmware incompatibility detected".After restart system info message: "firmware compatibility restored".Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi-500-01145, serial/lot #: 4.2.0, ubd: , udi#:.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|