| Model Number |
NIM4CM01 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that intra-op, the nim vital was giving an audible alarm approximately every minute with no corresponding alarm message displayed on screen.All connections were confirmed correct.The stimulation delivery tone was absent, although visual stimulation feedback was displayed under the stimulation current and muscle twitch was observed.No passive emg activity was detected with event capture turned off, and no nerve response was observed when stimulating the rln.Changing patient interfaces, using hard wire connections and rebooting the unit (unplugged for 30 seconds) did not resolve the issue.Adjusting the volume reportedly restored operation however, due to the abnormal behavior, medtronic recommended that the unit be assessed by a service engineer.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the codes fdm b17, fdr c20, fdc d16 and img g02030 are applicable for nim console.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: h6: the codes fdm b17, fdr c20, fdc d16 and img g02005 are applicable for nim patient interface.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|