• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT Puerto Rico Operations Co, Med Rel CD HORIZON® SOLERA® VOYAGER¿ Spinal System; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT Puerto Rico Operations Co, Med Rel CD HORIZON® SOLERA® VOYAGER¿ Spinal System; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 55850016540
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal therapy.It was reported that the set screw had difficulty passing through the upper inner threads of the tulip, and the operation time was extended.Several set screws were attempted.The screw was explanted and replaced.There were no patient symptoms or complications have been reported as a result of this event.Additional information received from the manufacturer representative that the procedure involved was transforaminal lumbar interbody fusion to l4/5.Additional information received from the manufacturer representative that the delay occurred is less than 60 minutes.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SOLERA® VOYAGER¿ Spinal System
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MDT Puerto Rico Operations Co, Med Rel
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT Puerto Rico Operations Co, Med Rel
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
+1(763)526-5677
MDR Report Key50928257
Report Number2647346-2026-00553
Device Sequence Number10759632
Product Code NKB
UDI-Device Identifier00643169768260
UDI-Public(01)00643169768260(11)240516(10)H5957499
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K170679
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55850016540
Device Catalogue Number55850016540
Device Lot NumberH5957499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured05/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-