Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 220V; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-3013-55 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during a routine check the cs100 intra-aortic balloon pump (iabp) had an external helium leak from the cylinder.No patient was involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.It was reported that helium gas manifold washer was damaged.The washer was replaced from the customer's available stock.After replacing the washer, fse connected the helium gas supply to the system and carried out all necessary safety checks.All safety parameters were found ok, and the valve leakage test was also completed successfully.The system was then checked with patient support, and everything was found working properly.After confirming normal operation, the system was handed over to the department in good working condition.No parts for return reported.No fault logs are available.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: event site name: (b)(6).Based on the fse's investigation, the helium leak was confirmed to be caused by a damaged helium gas manifold washer.The washer was replaced using the customer's available stock.Following replacement, all necessary safety checks and the valve leakage test were successfully completed.The system was checked with patient support and returned to the department in good working condition.The most probable cause was component reliability.This issue continues to be monitored within the manufacturer¿s quality system.
|
| |
|
Search Alerts/Recalls
|
|
|