| Catalog Number |
CB1413535040OTW |
| Medical Device Problem Codes |
Difficult to Insert (1316); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during treatment of a 90% stenotic, 30 mm plaque lesion in the mid right tibial/popliteal trunk peripheral artery, with minimal tortuosity and a 4 mm vessel diameter, a 3.5 x 40 chocolate.014 percutaneous transluminal angioplasty balloon catheter was used through a 6f destination sheath.The vessel was pre-dilated with a 3 x 80 nanocross percutaneous transluminal angioplasty balloon catheter, and the lesion remained visible on fluoroscopy.After the chocolate balloon catheter was opened, inspected, and flushed, resistance was felt when backloading a.014 command guidewire through the catheter.The guidewire was removed, a new.014 command es guidewire was opened, and the chocolate catheter was re-flushed and backloaded without issue.The chocolate catheter was tracked through the lesion and inflated without issue using 50/50 contrast.The catheter was removed for follow-up dsa imaging, and the lesion remained visible.Extreme resistance was encountered when reloading the chocolate catheter over the.014 command guidewire.Difficulty removing the chocolate catheter and flushing the balloon catheter lumen were reported.The chocolate catheter and gui dewire were removed safely, a.035 glidewire advantage was advanced, and the lesion was post-dilated with a 4 x 60 evercross percutaneous transluminal angioplasty balloon catheter without additional issues.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis the returned device was flushed, (photo 3) and a 0.014¿ guidewire was loaded without any difficulties, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|