| Model Number |
500004500000 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank tower, an error indicated as medication was available.The customer stated that this malfunction occurred while dispensing the hydromorphone medication.There were no adverse events or injuries reported due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 30-apr-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, a technical support specialist (tss) reviewed the reported issue and was unable to reproduce the problem during troubleshooting.The tss verified that medication dispensing was functioning correctly and confirmed that the customer was able to issue medication successfully.As the issue could not be replicated and normal operation was confirmed and confirmed that there were no issues with the device, and no further investigation was required.The system functioned as intended after the technical support specialist investigated the issue.
|
| |
|
Search Alerts/Recalls
|