| Model Number |
0998-00-0800-53 |
| Medical Device Problem Codes |
Pumping Problem (3016); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during therapy the cardiosave intra-aortic balloon pump (iabp) stopped pumping.A patient was involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse arrived on site and analyzed device logs.Fault code # 106 (communication error in the software data center) was recorded several times.The fse was unable to replicate any issue when testing on a patient simulator and test catheter.Executive processor pcba was replaced as a precaution.B.17 software was reinstalled, 30psi and helium transducers were recalibrated.Product passed all functional and safety tests per manufacturer specifications.The replaced executive processor pcba will be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - event site name: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|