A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 16-jan-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of this incident, the technical support specialist (tss) witnessed the customer select the patient, add the item to the patient order list, and click "issue item(s)." although a latch sound was heard, no drawer appeared to open.Further investigation showed the item screen showed that the medication was assigned to door 1.The customer was instructed to check the corresponding door and confirmed it was unlocked.The customer was then able to access and retrieve the medication, resolving the issue.No further assistance was required.The system functioned as intended after the technical support specialist assessed the device.
|