Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event:
model number/catalog number: sc-2317-70
serial number: 5070318
batch/lot number: 5070318
model/catalog description: infinion cx lead kit, 70cm
unique identifier (udi) #: (01)08714729861638(17)200626(10)5070318(21)5070318
model number/catalog number: sc-2317-70
serial number: 5072038
batch/lot number: 5072038
model/catalog description: infinion cx lead kit, 70cm
unique identifier (udi) #: (01)08714729861638(17)200704(10)5072038(21)5072038
investigation results, media review, device analysis
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.-a review was not completed of the device history records (dhr) for the device.The dhr review cannot be performed as no batch number was reported and a ship history cannot be performed since the upn was not reported.Labeling review
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review
a risk review was conducted and confirmed that the reported events are recognized within the products risk management documentation.These event types have been previously evaluated and are considered within the acceptable benefit-risk profile of the device.Investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|