BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Spectra WaveWriter?; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief Back to Search Results
Model Number SC-1160
Medical Device Problem Code Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 07/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patients implantable pulse generator (ipg) was experiencing loss of stimulation and pain the leads had high impedances.The patient underwent an implantable pulse generator (ipg) replacement procedure and was doing well postoperatively.The explanted device was not returned.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event: model number/catalog number: sc-2317-70 serial number: 5070318 batch/lot number: 5070318 model/catalog description: infinion cx lead kit, 70cm unique identifier (udi) #: (01)08714729861638(17)200626(10)5070318(21)5070318 model number/catalog number: sc-2317-70 serial number: 5072038 batch/lot number: 5072038 model/catalog description: infinion cx lead kit, 70cm unique identifier (udi) #: (01)08714729861638(17)200704(10)5072038(21)5072038 investigation results, media review, device analysis the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.-a review was not completed of the device history records (dhr) for the device.The dhr review cannot be performed as no batch number was reported and a ship history cannot be performed since the upn was not reported.Labeling review review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review a risk review was conducted and confirmed that the reported events are recognized within the products risk management documentation.These event types have been previously evaluated and are considered within the acceptable benefit-risk profile of the device.Investigation conclusion based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Spectra WaveWriter?
Common Device Name
Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50929783
Report Number3006630150-2026-07177
Device Sequence Number21514086
Product Code LGW
UDI-Device Identifier08714729951254
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/26/2020
Device Model NumberSC-1160
Device Catalogue NumberSC-1160
Device Lot Number338501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured06/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-