Datascope Corp. CARDIOSAVE HYBRID, TYPE B PLUG; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-0800-53 |
| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during use the cardiosave intra aortic balloon pump (iabp) batteries stopped charging.After this event occurred, this iabp unit was continued to be used throughout iabp therapy for this patient.As this iabp unit was connected to ac power source, this event did not affect continuing iabp therapy.There was no patient harm.A getinge field service engineer (fse) evaluated the unit and found that the power management board, which controls power and charging, was faulty.Therefore, the power management board was replaced.After the replacement, multiple discharge/charge cycles were repeated, and charging was confirmed to be normal.After the work was completed, an inspection was conducted based on the service manual, with good results.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to characterization limit e1:
initial reporter name: (b)(6).A getinge field service engineer (fse) evaluated the unit and found that the power management board, which controls power and charging, was faulty.Therefore, the power management board was replaced.After the replacement, multiple discharge/charge cycles were repeated, and charging was confirmed to be normal.After the work was completed, an inspection was conducted based on the service manual, with good results.The failure analysis and testing dept.Received part with a reported unit failure of the batteries stopped charging.The fat performed a visual inspection and found the part to have a blown q27 component.See attachment.This is a known failure and is due to a design issue.Retaining the part in the failure analysis and testing department per procedure.Rc1 ¿ there is no surge protection in the interface between the charging circuitry and the batteries installed in the power slots of the unit.Rc2 ¿ the tantalum capacitors used on the following two (2) pcbas are not within the capacitor manufacturer¿s (vishay) de-rating guidelines which may result in capacitor failure causing an overcurrent condition on the vbulk power supply which damages the power management board.
|
| |
|
Search Alerts/Recalls
|
|
|