| Model Number |
500001002502 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es auxiliary tower, the latch was double clicking and continuously failed.The customer stated that this malfunction occurred while dispensing the medication.The user was unable to access, which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 08-nov-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the door latch was double-clicking and continuously failing.A field service engineer (fse) cleaned the mini switch.Following the cleaning, no clicking was observed, and the issue was resolved.The system functioned as intended after the field service engineer repaired the device.
|
| |
|