| Model Number |
SC-1216 |
| Medical Device Problem Codes |
Positioning Problem (3009); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's implantable pulse generator (ipg) was sitting perpendicular to the surface of the skin causing charging difficulties.The patient underwent an ipg replacement procedure and was doing well postoperatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed that this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion:
the device was not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to the complaint.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code cause not established will be used.
|
| |
|