Hamilton medial ag received the following complaint description (summary):
it was reported that, during transport of an intubated patient to (b)(6) on (b)(6)2026, a hamilton-t1 (product no.161006, (b)(6), software version not reported) generated an alarm indicating that the oxygen supply was empty approximately 20 minutes before reaching pori.According to the information provided, the transport had started with a full 10-liter oxygen cylinder.The patient was connected to the ventilator without reported difficulty, and no warnings or abnormal device behavior were reportedly observed before the oxygen supply alarm occurred.The transport stopped at a hospital, where the primary oxygen cylinder was replaced with an available spare cylinder.Ventilation was continued using the same hamilton-t1.The complaint records medical intervention as ¿yes¿ and states that ventilation during transport was completed using the spare oxygen cylinder.No patient, user or third-party harm or adverse clinical outcome was reported.The user reported hearing a leak near the plastic retaining nut securing the return hose.According to the user, the expiratory valve had not been fully tightened, and the reported leakage stopped after the expiratory valve was tightened.It was reported that the user or hospital submitted a report to the competent authority.The authority report reference number is currently unknown.
|