| Model Number |
SC-1232 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient felt a zap near the implantable pulse generator (ipg) site.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results:
the ipg was not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
a labeling review was conducted and determined that the reported event is a known risk of implanting a pulse generator as part of a system to deliver spinal cord stimulation (scs).The event is consistent with risks disclosed in the device labeling.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Search Alerts/Recalls
|