| Model Number |
500001002506 |
| Medical Device Problem Code |
Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es tower was approximately 10 minutes behind the correct time, causing orders to not cross over in a timely manner.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 15-may-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device's was approximately 10 minutes behind the correct time, which was believed to be causing delays in order crossover.A technical support specialist (tss) corrected the device time and verified the update.The customer later confirmed that the issue was resolved and that no further assistance was required.The system functioned as intended after the technical support specialist troubleshot the device.
|
| |
|