• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Tandem Diabetes Care Tandem Mobi Insulin Pump with Control-IQ Technology; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tandem Diabetes Care Tandem Mobi Insulin Pump with Control-IQ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number 1004000
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pump displayed malfunction alarm code 12, which could be reset but alarm recurred after reset.There was no adverse impact to the customer¿s blood glucose level.Customer switched to multiple daily injections as backup insulin therapy, received guidance to reset pump, and was sent replacement pump with updated software; customer was instructed to put pump into storage mode, return problematic pump within 10 days using provided materials, and then program pump settings and reconnect continuous glucose monitor on replacement pump.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown.Mobile os: ios / ios_iphone 12 / ios_16.1.1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Tandem Mobi Insulin Pump with Control-IQ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50930837
Report Number3013756811-2026-176626
Device Sequence Number9508927
Product Code QFG
UDI-Device Identifier00389152370019
UDI-Public(01)00389152370019(22)1013701(11)240605
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K233044
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1004000
Device Catalogue Number1013700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age18 YR
Patient SexUnknown
-
-