• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. CARDIOSAVE HYBRID W/ E/F PLUG; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. CARDIOSAVE HYBRID W/ E/F PLUG; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-0800-55
Medical Device Problem Codes Battery Problem (2885); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50931132
Report Number2249723-2026-0004484
Device Sequence Number9734352
Product Code DSP
UDI-Device Identifier10607567108414
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K112372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2012
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-