• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Roche Diagnostics Creatinine plus ver.2; Creatinine test system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Roche Diagnostics Creatinine plus ver.2; Creatinine test system Back to Search Results
Catalog Number 08057524190
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable creatinine plus ver.2 results for several patient samples on a cobas c (b)(4) analytical unit.An interference with monoclonal gammopathy igm was suspected.Three patient examples were provided.Patient 1: the initial results were 337 µmol/l (serum) and 314 µmol/l (heparin plasma).The sample was tested in another laboratory on another module (cobas c 503), and the result was approximately 60 µmol/l the sample was repeated on the same module as the initial results, and the result was 300 µmol/l.Patient 2: the initial results were 338 µmol/l and 359 µmol/l (serum and heparin plasma).The results were obtained on 2 different cobas c 703 modules.A new sample was tested at a different laboratory on another module (cobas c 501), and the result was 87 µmol/l.A third sample was tested on one of the same modules as the initial sample, and the result was approximately 330 µmol/l.Patient 3: the initial result was 215 µmol/l.It was alleged that a new sample was tested at another laboratory, but the result was not provided.
 
Additional Manufacturer Narrative
Only one of the cobas c 703 module serial numbers was provided.The serial number was (b)(6).The investigation is ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Creatinine plus ver.2
Common Device Name
Creatinine test system
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50931151
Report Number1823260-2026-03996
Device Sequence Number10759200
Product Code JFY
UDI-Device Identifier07613336121160
UDI-Public07613336121160
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K024098
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08057524190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-