| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a battery charger was unable to charge a battery pack.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery charger has been completed.The reported problem (unable to charge batteries) has been confirmed.Upon investigation the battery charger was unable to charge/recognize a battery pack.The cause for the failure was contamination on the battery board pca.The root cause for the contamination was ingress of an unknown liquid.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.There was no adverse event that resulted from the damaged charger.
|
| |
|