MEDTRONIC PUERTO RICO OPERATIONS CO. InterStim X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
| Model Number |
97800 |
| Medical Device Problem Codes |
Break (1069); Failure to Deliver Energy (1211); Energy Output Problem (1431); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from a manufacturer representative (rep) regarding a patient with an implanted neurostimulator (ins) for urinary dysfunction (incontinence, urgency, and/or frequency).The rep reported a 97800 issue.The lead was fractured, damaged, or broken.As a result, there was loss of stimulation.The rep stated that the programmer wouldn't allow stimulation to increase.A replacement was scheduled and they found the lead was broken at the insertion of the battery header with no obvious twiddling of the device.The damage was caused or discovered during surgery.There were impedances on all 4 electrodes prior to the case.The cause was not known.The device was replaced on (b)(6) 2026 and the issue was resolved.The rep stated that they had possession of the device and that it would be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 978b128 lot# (b)(6) serial# implanted: (b)(6) 2025-explanted: (b)(6) 2026 product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 978b128, serial/lot #: (b)(6), ubd: 22-jul-2026, udi#:(b)(4) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|