| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a patient's electrode belt could not run detect and treat testing.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (cannot run detect and treat testing) was confirmed.The belt was unable to pass incoming testing.The pgnd wire was reading open in the trunk cable.The root cause for open wire was unable to be identified.There was no adverse event that resulted from the damaged belt.
|
| |
|